The BEP-M600L Benchtop Multi-parameter Analyzer adopts a multi-purpose measurement method, which is suitable for the fields of health disease control, environmental protection, biological agriculture and forestry, water quality analysis, scientific research and institutions of higher learning.
General Features
pH
Ion
Conductivity
DO
| Model | BEP-M600L | |
| Parameters | pH/EC/ISE/DO/Temp. (mV/ORP/pX/Resistivity/TDS/Sal./DO Saturation) | |
| pH | Range | -2.000 to 20.000 pH |
| Resolution | 0.1, 0.01, 0.001pH | |
| Accuracy | ±0.002 pH | |
| Calibration Points | Up to 6 | |
| Standard Customization | Yes | |
| Calibration Reminder | Yes | |
| Standard Recognition | NIST, GB, DIN buffers | |
| Slope Limit | Yes | |
| mV/ORP | Range | -2000.00 to 2000.00 mV |
| Resolution | 0.1, 0.01 mV | |
| Accuracy | ±0.03% FS | |
| EH ORP Mode | Yes | |
| ORP Calibration Points | 1~2 custom point (relative mV) | |
| pX | Range | -2.000 to 20.000 pX |
| Resolution | 0.1, 0.01, 0.001 pX | |
| Accuracy | ±0.002 pX | |
| Calibration Points | Up to 6 | |
| ISE | Range | 1.000E-9 to 9.999E9 |
| Units | mol/L, mmol/L, g/L, mg/L, μg/L | |
| Resolution | Up to 4 significant digits | |
| Accuracy | ±0.3% | |
| Calibration Points | Up to 6 | |
| Conductivity | Range | 0.000 μS/cm to 3000 mS/cm |
| Resolution | 0.001 μS/cm minimum, various with range selection | |
| Accuracy | ±0.5% FS | |
| Reference Temperature | 5, 10, 15, 18, 20, 25 ºC | |
| Calibration Points | Up to 5 | |
| Standard Recognition | 146.5μS/cm, 1408μS/cm, 12.85mS/cm, 111.3mS/cm | |
| Resistivity | Range | 5.00 Ω·cm~100.00 MΩ·cm |
| Resolution | 0.01 Ω·cm minimum | |
| Accuracy | ±0.5% FS | |
| TDS | Range | 0.000 mg/L~1000 g/L |
| Resolution | 0.001ppm minimum, changed with range | |
| Accuracy | ±0.5%FS | |
| Salinity | Type | NaCl% |
| Range | (0.00~8.00)% | |
| Resolution | 0.01% | |
| Accuracy | ±0.1% | |
| Dissolved | Range | 0.00 to 99.99 mg/L |
| Oxygen Concentration – Polarographic | Resolution | 0.01 mg/L |
| Accuracy | ±0.1 mg/L | |
| Calibration Reminder | Yes | |
| Calibration Points | Air-saturated water or zero point | |
| Barometric Compensation | Yes | |
| Manual Salinity Factor Correction | Yes | |
| % Saturation -Polarographic | Range | (0.0 to 600)% |
| Resolution | 0.10% | |
| Accuracy | ±2.0% | |
| Temperature | Range | -5 to 130 ºC |
| Unit | ºC | |
| Resolution | 0.1 | |
| Accuracy | ±0.1 | |
| Measurement | Reading Mode | Auto Read (Fast, Medium, Slow), Timed, Continuous |
| Reading Prompts | Reading, Stable, Locked | |
| Temp. Compensation | ATC, MTC | |
| Data management | Data Storage | 1000 Groups |
| GLP Features | Yes | |
| Log Management | Yes | |
| Inputs | pH Electrode | BNC(Q9) |
| DO with Temp. Probe | 6-pin Mini-DIN | |
| Conductivity with Temp. Probe | 8-pin Mini-DIN | |
| Outputs | USB, RS 232 | USB 2.0 flash memory device, printer, PC |
| Display options | Backlight | Yes |
| Auto Shutdown | 1~60 min, off | |
| IP Rating | IP54 | |
| Date and Time | Yes | |
| General | Electricity | AC Adapter, 100-240V AC input, DC24V output |
| Dimensions | 280×280×130 mm | |
| Weight | 2500g | |