Детектор: FID, TCD, ECD и FPD являются дополнительными. Стандартная двухканальная карта сбора данных. Функция самопроверки и функция автоматической идентификации неисправностей. Автоматическое управление. Разделение потока/коэффициент разделения.
1.The host machine adopts a 7-inch color touch screen with friendly man-machine interface.
2. Computer back control and host touch screen to achieve synchronous bidirectional control.
3. The multi-core, 32-bit embedded hardware system ensures the reliable operation of the instrument.
4. Carrier/hydrogen/air flow (pressure) digital display.
5. Gas shortage alarm protection function; Heating control protection function (when opening the door of the column box, the motor of the column box fan and the heating system will shut down automatically).
6. Split flow/split ratio can be automatically controlled to save carrier gas.
7. Configure automatic sampler installation and positioning interface to match automatic sampler of various specifications.
8. Data acquisition is a standard dual-channel data acquisition card with a sampling time of 50ms.
9. Using logarithmic amplification plate, detection signal no cut-off value, synchronous external trigger function, can be started by external signals (automatic sampler, thermal analyzer, etc.) at the same time the host and workstation.
10. It has perfect system self-check function and fault automatic identification function.
11. With eight external event extension function interface, can be selected with a variety of control valves, and according to their own set time sequence work.
12. The external link mode is network port connection (RJ45), which provides convenience for remote control of data.
Column temperature box
Sampler
Detector
High voltage switch control
Baseline signal display
Ignition coil control
Detection limit: ≤ 3×10-12 g/s
Sample: N-hexadecane (minimum detection quantity: 3pg/s)
Dynamic range: 107
Bridge voltage switch control
Bridge current setting: 0 ~ 220 mA
Sensitivity: 5000 mV×mL/mg
Dynamic range: 105
Radiation source: Ni63
Detection limit: ≤8×10-14 g/s
Sample: R 666 (minimum detection quantity: 80FG/s)
Dynamic range: 103
Temperature setting: Maximum 350℃
Detection limit: ≤2×10-12 g/s (P) ≤4×10-11 g/s (s)
Sample: Methyl parathion
Dynamic range: P 103
Dynamic range: S 102
Heating area